ZiP-P2 Instrument
Product code: P002736
​
​
The ZiP-P2 instrument operates the ZiP-P2 cartridge and provides high sensitivity molecular (nucleic acid amplification) diagnostic results. The instrument is rugged and portable. It is suitable for use in professional point-of-care settings, field or pen-side applications and laboratories. The instrument provides a screen-prompted workflow that includes result interpretation and automated reporting. It supports connection of a barcode reader and a label printer. Network options include server connection and full LIS.
​
Research Use Only.
Always follow the user manual instructions.
​
*This product is not available for purchase by the general public.
The ZiP-P2 instrument is supplied with a power supply and region-specific power adapters. Accessories may be purchased separately through ZiP Diagnostics. Contact ZiP Sales team for more information.
​
Manufactured in Australia.
​
The ZiP-P2 instrument has the following specifications:
Cartridge Operation
Up to 2 ZiP-P2 cartridges simultaneously, random access
Data Storage
Up to 300 test results can be stored on the instrument
History Records allow search and retrieval
Archive or export via USB / Ethernet to FTP or HL7 systems
Test Duration
Up to 30 minutes
Colour Touch Screen
109 mm (4.3") LCD display capacitive touch screen.
Simple workflow management
Ease of use, colour icon driver menus and prompts
Measurement Technology
Fluorescence detection
Communications
USB port for data export, software update and test types import
Barcode reader port
Label printer port
Ethernet RJ45 port for network connectivity
LIS/HL7 support
Test Result
PDF data file or print to label printer
Power
12 V DC
3.0 Amps
Power supply adaptor is configured for different regions (Australia, UK, US, and Europe)
Dimensions
Width 185 mm (7.28")
Depth 205 mm (8.07")
Height 122 mm (4.80")
Weight
2.3 kg (5.1 lb)
Design Lifetime
5 years or 10,000 tests, whichever occurs first
Operating Environment
Indoor use
10°C to 30°C, 20% to 80% RH (non-condensing)
0 to 2000 m altitude
Storage Environment
2°C to 45°C, 20% to 80% RH (non-condensing)
Standards Compliance
EMC Directive: IEC 61326-1 - IEC 61326-2-6
IEC 61010-1; 61010-2-010; 61010-2-101,
IEC 62471
RoHS 2 & Directive 2015/863/EU
WEEE Directive 2
REACH, EC 1907/2006 (as per Article 33, a list of "Substances of Very High Concern" identified within articles of the product in concentration above 0.1.% w/w is available on request)
Warranty
1 year